SENSe Subsystem:FUNCtion EM510
5.81 4065.7763.32-01.00
. . [:ON]?
Query about sensor functions being switched on.
Parameters:
none
Result:
List of sensor functions switched on. If no function is active, a zero string ("") is output. The list has a
specific order:
1. level measurement function
2. offset measurement function
3. reserved
4. AM modulation index measurement function
5. AM positive modulation index measurement function
6. AM negative modulation index measurement function
7. FM frequency deviation measurement function
8. FM positive frequency deviation measurement function
9. FM negative frequency deviation measurement function
10. PM phase deviation measurement function
11. bandwidth measurement function
The following strings are to be expected:
"VOLT:AC" level measurement switched on
"FREQ:OFFS" offset measurement switched on
"FSTR" field strength measurement switched on
"AM" AM modulation index measurement switched on
"AM:POS" AM positive modulation index measurement switched on
"AM:NEG" AM negative modulation index measurement switched on
"FM" FM frequency deviation measurement switched on
"FM:POS" FM positive frequency deviation measurement switched on
"FM:NEG" FM negative frequency deviation measurement switched on
"PM" PM phase deviation measurement switched on
"BAND" bandwidth measurement switched on
Example:
FUNCtion? -> "VOLT:AC", "FREQ:OFFS"
. . . :COUNt?
Query about the number of sensor functions being active.
Parameters:
none
Result:
Number of sensor functions being active
Example:
FUNCtion:Count? -> 2