EM510 SENSe Subsystem:GCONtrol
4065.7763.32-01.00 5.82
. :GCONtrol
. . [:FIXed|MGC] <numeric_value>|UP|DOWN|MINimum|MAXimum
Setting the MGC value.
Parameters:
<numeric_value> gain control factor in dBµV
-30 dBµV = no gain control -> maximum sensitivity
110 dBµV = maximum gain control -> minimum sensitivity
UP|DOWN increase|decrease of the MGC value by the value set in the
command
SENSe:GCONtrol[:FIXed|MGC]:STEP[:INCRement].
MINimum|MAXimum sets the smallest|largest MGC value
*RST state:
50 dBuV
Example:
GCONtrol 50
. . [:FIXed|MGC]? [MINimum|MAXimum]
Query about the MGC value.
Parameters:
none query about the current MGC value
MINimum|MAXimum query about the smallest|largest MGC value
Result:
Gain control
Example:
GCONtrol? -> 50
. . :AUTO:TIME SLOW | DEFault | FAST
. . :AGC:TIME SLOW | DEFault | FAST
Setting the gain control time.
Parameters:
SLOW slow gain control time
DEFault default gain control time
FAST fast gain control time
*RST state:
DEF
Example:
GCONtrol:AUTO:TIME FAST