EM510 SENSe Subsystem :SWEep
4065.7763.32-01.00 5.96
. . :DWELl <numeric_value>|MINimum|MAXimum|INFinity
Setting the dwell time with the hold criterion fulfilled (T-DWELL in FSCAN-CONFIG menu).
Note:
According to the SCPI standard, this command is used to set the dwell time per scan step, i.e. the time
required by a step. This definition is met in the EM510 if the squelch is switched off. The hold criterion is
then fulfilled for each step.
Parameters:
<numeric_value> dwell time in seconds
MINimum|MAXimum lowest|highest dwell time
INFinity infinite dwell time
*RST state:
0.5 s
Example:
SWEep:DWELl 10 ms
. . :DWELl? [MINimum|MAXimum]
Query about the dwell time with hold criterion fulfilled.
Parameters:
none query about the current dwell time
MINimum|MAXimum query about the lowest|highest dwell time
Result:
Dwell time in seconds; ; 9.9E37 is output for an infinite number.
Example:
SWEep:DWELl? 0.010
. . :HOLD
. . . :TIME <numeric_value>|MINimum|MAXimum
Setting the hold time during signal-controlled scan continuation (T_NOSIG in FSCAN-CONFIG menu).
If the signal disappears during the dwell time, the hold time is started. After completion of the hold time,
the scan is continued with the next frequency even if the dwell time has not yet been completed. If the
signal during the hold time exceeds the squelch threshold, the hold time is reset and the end of the
dwell time or the renewed disappearance of the signal is awaited. The hold time is only important if the
control function "STOP:SIGNal" (see SENSe:SWEep:CONTrol) is switched on.
Parameters:
<numeric_value> hold time in seconds
MINimum|MAXimum lowest|highest hold time
*RST state:
0 s
Example:
SWEep:HOLD:TIME 10 ms