EasyManua.ls Logo

R&S EM510 - STATus:QUEStionable and TRACe Registers

R&S EM510
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control via LAN EM510
5.137 4065.7763.32-01.00
5.5.3.6 STATus:QUEStionable Register
This register contains information on ambiguous device states. They can occur, for example if the
device is operated outside its specification range. It can be queried using the commands
STATus:QUEStionable:CONDition? or STATus:QUEStionable[:EVENt]?.
Table 5-5: Bit allocation of STATus:QUEStionable register
Bit No. Meaning
0 VOLTage
This bit is set if an ambiguous supply voltage occurs. Due to that all test points of the supply voltage are checked.
4
TEMPerature
This bit is set if the internal temperature is too high. All temperature test points in the equipment are tested.
5 FREQuency
This bit is set if an internal oscillator frequency is ambiguous. Due to that the test points of the oscillators are
checked.
9 OVERload
This bit is set if the IF section is overdriven by too high an input level. Then the result of a level measurement is
questionable.

Table of Contents