Installation and Cabling EM510
3.3 4065.7763.32-01.00
3.5.1 Connection of Power Supply
The EM510 is suitable for AC mains operation as well as DC operation.
Connection of the EM510 both to the mains and the DC source permits AC/DC switch-over, i.e. if there
is a power failure the DC source (battery) will take over automatically.
3.5.1.1 Mains Connection
Attention:
Make sure that the available mains voltage is between 110 and 240 VAC.
The EM510 is connected to the mains voltage by the plug with mains filter on the rear panel.
The power cable is supplied as an accessory.
3.5.1.2 DC Source Connection
Attention:
Make sure that the available supply voltage is between 12 and 32 V.
Observe correct voltage polarity when connecting.
Incorrect polarity may blow the rear panel fuse or damage the EM510.
Note: If MIL-Spec regarding EMC must be fulfilled, a screened battery cable not exceeding 3 m in
length is necessary. It is delivered by Rohde & Schwarz on request.
The EM510 is connected to an external 12 to 32-VDC source (e.g. battery) by the connector X40 12 TO
32 VDC on the rear panel.
Recommended connector: NeutrikSpeakonNL4FX