EasyManua.ls Logo

R&S EM510 - CALibration Subsystem: Oscillator Calibration

R&S EM510
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CALibration Subsystem EM510
5.21 4065.7763.32-01.00
5.3.5 CALibration Subsystem
. : ROSCillator[:DATA] <numeric_value> [MINimum|MAXimum|UP|DOWN]
Changing the calibration value (D/A-converter value) for setting the exact OCXO reference frequency.
Parameter:
<numeric_value> 0 ... 4095
*RST state:
none
default state (not calibrated)
2048
Example:
CALibration:ROSCillator UP
. : ROSCillator[:DATA]? [MINimum|MAXimum]
Query about the calibration value (D/A-converter value) for setting the exact OCXO reference
frequency.
Parameter:
none query about the current calibration value
MINimum|MAXimum query about the lowest/highest calibration value
Result:
Calibration value or D/A-converter value for setting the exact OCXO reference frequency.
Example:
CALibration:ROSCillator? MAX -> 4095

Table of Contents

Related product manuals