4065.7763.32-01.00 5.38
5.3.11 MEASure Subsystem
MEASure
. :BANDwidth:MODE XDB|BETA
Selection of bandwidth measurement method.
Parameters:
XDB bandwidth measurement method x dB
BETA bandwidth measurement method Beta %
*RST state:
XDB
Example:
MEASure:BANDwidth:MODE XDB
. :BANDwidth:MODE?
Query about the selected bandwidth measurement method.
Result:
XDB bandwidth measurement method x dB
BETA bandwidth measurement method Beta %
Example:
MEASure:BANDwidth:MODE? -> XDB
. :BANDwidth:XDB<numeric_value>
Configuration of bandwidth measurement method x dB.
Parameter:
<numeric_value> 0.0 to 100.0 dB
*RST state:
26.0 dB
Example:
MEASure:BANDwidth:XDB 31
. :BANDwidth:XDB?
Query about the configuration of bandwidth measurement method x dB.
Result:
dB value of bandwidth measurement method x dB.
Example:
MEASure:BANDwidth:XDB? -> 31.0