EM510 SENSe Subsystem :DATA?
4065.7763.32-01.00 5.62
Result:
Level in dBuV, offset in Hz
The output format will be generated with the command FORMat:DATA according to the setting:
ASCii normal ASCII output
PACKed <Definite Length Block>:
level in 1/10 dBuV (2 Byte)
offset in Hz (4 Byte)
AM-modulation index in 1/10 % (2 bytes),
AM-positive modulation index in 1/10 % (2 bytes),
AM-negative modulation index in 1/10 % (2 bytes),
FM frequency deviation in Hz (4 bytes),
positive frequency deviation in Hz (4 bytes),
negative frequency deviation in Hz (4 bytes),
phase deviation in 1/100 rad (2 bytes),
bandwidth in Hz (4 bytes)
Examples:
SENSe:DATA? -> 23.4, -2500
SENSe:DATA? "VOLT:AC" -> 23.4
SENSe:DATA? "FREQuency:OFFSet" -> -2500