Operating Manual 1302.6163.12 - 03 4.125
R&S ESU Instrument Functions
Analyzer Mode
Fig. 4.3 Pulsed signal GATE OFF
Fig. 4.4 TDMA signal with GATE ON
The gated-sweep mode is activated by the GATED TRIGGER softkey. The setting of
the mode takes place in the GATE SETTINGS submenu.
GATED
TRIGGER
The GATED TRIGGER softkey switches the sweep mode with gate on and off.
When gate is switched on, a gate signal applied to the rear panel connector EXT
TRIGGER/GATE or the internal IF power detector controls the sweep of the
R&S ESU. This selection is made via the EXTERN and IF POWER softkeys for trig-