R&S ESU Instrument Functions
Analyzer Mode
4.232 Operating Manual 1302.6163.12 - 03
Table 4-5 Attributes and child nodes of the Range element (normal ranges)
DefaultLimitFailMode Absolute | Relative |
Absolute and Relative |
Absolute or Relative
Yes
Child Node Attribute Value Parameter
Description
Mandatory
Child Node Attribute Value Parameter
Description
Mandatory
Index 0...19 Indexes are continous
and have to start with 0
Yes
Name <string> Name of the range Only if
ReferenceChannel
contains a name and
range is reference
range
ShortName <string> Short name of the range No
ChannelType TX | Adjacent Yes
WeightingFilter Only if
ReferencePower
method is "TX
Channel Power" and
the range is the
reference range"
Type RRC | CFILTer Type of weighting filter Yes
RollOfffactor 0...1 Excess filter bandwidth Only if filter type is
RRC
Bandwidth <bandwidth in Hz> Filter bandwidth Only if filter type is
RRC
FrequencyRange Yes
Start <frequency in Hz> Start value of the range Yes
Stop <frequency in Hz> Stop value of the range Yes
Limit A Range must contain
exactly two limit nodes;
one of the limit nodes
has to have a relative
unit (e.g. dBc), the other
one must have an
absolute unit (e.g. dBm)
Yes
Start Value <numeric value> Power limit at start
frequency
Yes
Unit dBm/Hz | dBm | dBc | dBr | dB Sets the unit of start
frequency
Yes
Stop Value <numeric value> Power limit at stop
frequency
Yes
Unit dBm/Hz | dBm | dBc | dBr | dB Sets the unit of stop
frequency
Yes
LimitFailMode Absolute | Relative | Absolute
and Relative | Absolute or
Relative
No (if specified, it has
to be equal to Default-
LimitFailMode