Operating Manual 1302.6163.12 - 03 4.237
R&S ESU Instrument Functions
Analyzer Mode
• Limit Check
Sets the type of limit check for all ranges.
ABSOLUTE:
Checks only the defined absolute limits.
RELATIVE:
Checks only the relative limits. Relative limits are defined as relative to the
measured power in the reference range.
ABS & REL:
Combines the absolute and relative limit. The limit check fails when both limits are
violated.
ABS or REL:
Combines the absolute and relative limit. The limit check fails when one of the
limits is violated.
Remote command: ESP:RANG1:LIM:STAT ABS | REL | AND | OR
• Abs Limit Start
Sets the absolute limit value at the start frequency of the range. The unit depends
on the unit selected via the UNIT softkey.
Remote command: ESP:RANG1:LIM:ABS:STAR 10
• Abs Limit Stop
Sets the absolute limit value at the stop frequency of the range. The unit depends
on the unit selected via the UNIT softkey.
Remote command: ESP:RANG1:LIM:ABS:STOP 20
• Rel Limit Start
Sets the relative limit value at the start frequency of the range. The unit is dBc.
Remote command: ESP:RANG1:LIM:REL:STAR -20
• Rel Limit Stop
Sets the relative limit value at the stop frequency of the range. The unit is dBc.
Remote command: ESP:RANG1:LIM:REL:STOP -20
INS BEFORE
RANGE
Inserts a range to the left of the current range / column of the sweep list.
Remote command: ESP:RANG1:INS:BEF
INS AFTER
RANGE
Inserts a range to the right of the current range / column of the sweep list.
Remote command: ESP:RANG1:INS:AFT
DELETE
RANGE
Deletes the current range / column from the sweep list. All ranges above the deleted
one are adjusted by one (e.g. Range 6 turns to Range 5).
Remote command: ESP:RANG:DEL