R&S ESU Instrument Functions
Analyzer Mode
4.238 Operating Manual 1302.6163.12 - 03
NEXT RANGES Scrolls through the sweep list in forward direction if more than 5 ranges are defined.
The first page shows Range 1 through 5, the second page shows Range 6 to 11 etc.
PREVIOUS
RANGES
Scrolls through the sweep list in backwards direction if more than 5 ranges are
defined. The first page shows Range 1 through 5, the second page shows Range 6
to 11 etc.
REFERENCE
RANGE
Opens a submenu to define additional parameters for the reference power measure-
ment.
POW REF
CHAN PEAK
Selects the type of power measurement in the reference range.
•Peak
Measures the highest peak within the reference range.
• Chan
Measures the channel power within the reference range (integral bandwidth
method).
Remote command: ESP:RTYP PEAK
TX
BANDWIDTH
Defines the bandwidth used for measuring the channel power. Possible values are
from 20 Hz to the span of the reference range.
Remote command: ESP:BWID 20 HZ
RRC FILT
(ON OFF)
Activates or deactivates the RRC filter.
Remote command: ESP:FILT:RRC:STAT ON
ALPHA BT Sets the rolloff factor of the RRC filter.
Remote command: ESP:FILT:RRC:ALPH 0.22
START MEAS Starts the Spectrum Emission Mask measurement. In Single Sweep mode, the
R&S ESU automatically stops the measurement.
Remote command: INIT:ESP
STOP MEAS Stops the Spectrum Emission Maskj measurement.
Remote command: ABOR
LIST EVAL Opens a submenu to edit the list evaluation settings.
POW REF CHAN PEAK
TX BANDWIDTH
RRC FILT (ON OFF)
ALPHA BT
LIST EVAL (ON OFF)
MARGIN