R&S ESU Instrument Functions
Basic Settings
4.282 Operating Manual 1302.6163.12 - 03
This softkey is only available, if a HP languge is selected via the GPIB LANGUAGE
softkey.
Remote command: SYST:RSW ON
COUPLING
FSP/HP
The COUPLING FSP/HP softkey controls the default coupling ratios for:
• span and resolution bandwidth (Span/RBW) and
• resolution bandwidth and video bandwidth (RBW/VBW)
for the HP emulation mode.
In case of FSP the standard parameter coupling of the instrument is used. As a
result in most cases a shorter sweeptime is used than in case of HP.
The softkey is only available in HP emulation mode.
Remote command: SYST:HPC FSP
User Port Configuration
The instrument provides a parallel interface, which is 8 bits wide. Over these ports,
arbitrary bit patterns can be output or input.
USER PORT
The USER PORT softkey opens the submenu for setting the direction of the data
transmission and for entering the values of the individual ports.
USER PORT IN/
OUT
The USER PORT IN/OUT determines in which direction the interface transmits data.
USER PORT (read operation)
The value is indicated in a window. A new readout is started by pressing READ.
Pressing OK closes the window
Remote command: INP:UPOR:STAT ON
INP:UPOR?
USER PORT OUT (write operation)
The bit pattern selected with softkeys PORT 0 to PORT 7 is output.
Remote command: OUTP:UPOR:STAT ON
PORT 0 0/1 to
PORT 7 0/1
The PORT 0 0/1 to PORT 7 0/1 softkeys determine the output value for port 1 to 7.
These softkeys are only available for setting USER PORT OUT.
Remote command: OUTP:UPOR STAT ON
OUTP:UPOR #B10110010
USER PORT IN/OUT
PORT 0 0/1 to PORT 7 0/1
00000000
USER PORT INPUT