R&S ESU Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.328 Operating Manual 1302.6163.12 - 03
Table 4-6 Measurement accuracy levels
Accuracy Enhancement
label
Reason/Limitation
High NOR No difference between reference setting and
measurement
Medium APX
(approximation)
Change of the following settings:
coupling (RBW, VBW, SWT)
reference level, RF attenuation
start or stop frequency
output level of tracking generator
frequency offset of tracking generator
detector (max. peak, min. peak, sample, etc.)
Change of frequency:
max. 625 points within the set sweep limits
(corresponds to a doubling of the span)
- Aborted
normalization
625 and more extrapolated points within the current
sweep limits (in case of span doubling)
At a reference level (REF LEVEL) of -10 dBm and at a tracking generator output
level of the same value the R&S ESU operates without overrange reserve, i.e. the
R&S ESU is in danger of being overloaded if a signal is applied whose amplitude
is higher than the reference line. In this case, either the message OVLD for over-
load is displayed in the status line or the display range is exceeded (clipping of the
trace at the upper diagram border = Overrange).
Overloading can be avoided as follows:
Reducing the output level of the tracking generator (SOURCE POWER,
NETWORK menu)
Increasing the reference level (REF LEVEL, AMPT menu)