Operating Manual 1302.6163.12 - 03 5.27
R&S ESU Remote Control – Basics
Status Reporting System
5.7.3.8 STATus:QUEStionable:LIMit<1|2> Register
This register comprises information about the observance of limit lines in the corre-
sponding measurement window (LIMit 1 corresponds to Screen A, LIMit 2 to Screen
B). It can be queried with commands STATus:QUEStionable:LIMit<1|2>:
CONDition? and STATus:QUEStionable:LIMit<1|2>[:EVENt]?.
Bit
No.
Meaning
0 LIMit 1 FAIL
This bit is set if limit line 1 is violated.
1 LIMit 2 FAIL
This bit is set if limit line 2 is violated.
2 LIMit 3 FAIL
This bit is set if limit line 3 is violated.
3 LIMit 4 FAIL
This bit is set if limit line 4 is violated.
4 LIMit 5 FAIL
This bit is set if limit line 5 is violated.
5 LIMit 6 FAIL
This bit is set if limit line 6 is violated.
6 LIMit 7 FAIL
This bit is set if limit line 7 is violated.
7 LIMit 8 FAIL
This bit is set if limit line 8 is violated.
8 to 14 not used
15 This bit is always 0.