R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
6.46 Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:MARKer<1...4>:SCOupled[:STATe] ON | OFF
This command switches the coupling of the receiver frequency settings from the corresponding
subscans to the marker frequency on or off.
Example: "CALC:MARK:SCO ON"
Characteristics: *RST value: ON
SCPI: device-specific
Mode: R
The numeric suffix in MARKer<1...4> is not significant.
CALCulate<1|2>:MARKer<1...4>:LOEXclude ON | OFF
This command switches the local oscillator suppression for peak search on or off. This setting is valid
for all markers and delta markers in all measurement windows. The numeric suffixes 1|2 and 1...4 are
irrelevant.
Example: "CALC:MARK:LOEX ON"
Characteristics: *RST value: ON
SCPI: device-specific
Mode: A-F
CALCulate<1|2>:MARKer<1...4>:Y?
This command queries the measured value of the selected marker in the indicated measurement
window. The corresponding marker is activated before or switched to marker mode, if necessary.
To obtain a valid query result, a complete sweep with synchronization to the sweep end must be
performed between the activation of the marker and the query of the y value. This is only possible in
single-sweep mode.
The query result is output in the unit determined with the CALCulate<1|2>:UNIT:POWer command.
Only with linear level scaling is the output in %.
Example: "INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK2 ON"
'Switches marker 2 in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK2:Y?"
'Outputs the measured value of marker 2 in screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode: A