R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
6.48 Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:MARKer<1...4>:MAXimum:RIGHt
This command positions the specified marker to the next smaller maximum value to the right of the
current value (i.e. in ascending X values) on the corresponding trace.
Example: "CALC:MARK2:MAX:RIGH"
'Positions marker 2 in screen A to the next lower maximum value to the right of the
current value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MAXimum:LEFT
This command positions the specified marker to the next smaller maximum value to the left of the
current value (i.e. in descending X values) on the trace.
Example: "CALC:MARK2:MAX:LEFT"
'Positions marker 2 in screen A to the next lower maximum value to the left of the
current value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MAXimum:AUTO ON | OFF
This command switches an automatic maximum peak search for marker 1 at the end of each particular
sweep on and off. The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,
EXCLUDE LO) are taken into account. The numeric suffix at MARKer<1...4> is irrelevant.
Example: "CALC:MARK:MAX:AUTO ON"
Activates the auto search function for marker 1.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
If no next smaller maximum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: -200) is produced.
If no next smaller maximum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: -200) is produced.