Operating Manual 1302.6163.12 - 03 6.133
R&S ESU Remote Control – Description of Commands
INPut Subsystem
INPut<1|2>:TYPE INPUT1 | INPUT2
The command selects the signal input. INPUT1 is RF input 1 and INPUT2 is RF input 2.
Example: "INP:TYPE INPUT1"
Characteristics: *RST value: INPUT1
SCPI: conform
Mode: R, A
INPut<1|2>:UPORt[:VALue]?
This command queries the control lines of the user ports.
Example: "INP:UPOR?"
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A
This command is a query and therefore has no *RST value.
INPut<1|2>:UPORt:STATe ON | OFF
This command toggles the control lines of the user ports between INPut and OUTPut.
Example: "INP:UPOR:STAT ON"
Characteristics: *RST value: ON
SCPI: device-specific
Mode: R, A
With ON, the user port is switched to INPut, with OFF to OUTPut.
INPut<1|2>:GAIN:LNA:STATe ON | OFF
This command switches the low noise preamplifier on and off (available with option 
R&S ESU-B24.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R