Operating Manual 1302.6163.12 - 03 6.171
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]ESPectrum:RANGe<1…20>:COUNt?
This command indicates the number of defined ranges of the Spectrum Emission Mask measurement.
The numeric suffixes <1|2> at SENSe and <1 to 20> at RANGe are irrelevant for this command.
Example: "ESP:RANG:COUN?"
'indicates the number of ranges
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
[SENSe<1|2>:]ESPectrum:RANGe<1…20>:DELete
This command deletes a range of the Spectrum Emission Mask measurement configuration.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example: "ESP:RANG2:DEL"
'deletes range 2
Characteristics: *RST value:-
SCPI: device-specific
Mode: A
[SENSe<1|2>:]ESPectrum:RANGe<1…20>:FILTer:TYPE NORMal | CFILter | RRC | P5 | P5Digital |
PULSe
This command sets the filter in the Spectrum Emission Mask measurement. For details on filter type
refer to “Filter Types” on page 4.117.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example: "ESP:RANG2:FILT:TYPE RRC"
'RRC filter
Characteristics: *RST value: NORMal
SCPI: device-specific
Mode: A
[SENSe<1|2>:]ESPectrum:RANGe<1…20>[:FREQuency]:STARt <numeric_value>
This command sets the relative start frequency of a range in the Spectrum Emission Mask
measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example: "ESP:RANG2:STAR -2.52MHz"
'start frequency of range 2 to -2.52 MHz
Characteristics: *RST value:-
SCPI: device-specific
Mode: A