Operating Manual 1302.6163.12 - 03 6.173
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:INSert AFTer | BEFore
This command inserts a new range before or after the specified range. The range numbers are
updated accordingly. The numeric suffix at RANGe<1...20> specify the range.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example: "ESP:RANG3:INS BEF"
'Inserts a new range before Range 3
"ESP:RANG1:INS AFT"
'Inserts a new range after Range 1
Characteristics: *RST value:-
SCPI: device-specific
Mode: A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:ABSolute:STARt <numeric_value>
This command defines the absolute limit at the start frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter: -400 to +400 dBm
Example: "ESP:RANG5:LIM:ABS:STAR -40"
'Absolute limit at start frequency in range 5 is -40 dBm
Characteristics: *RST value: -13dBm
SCPI: device-specific
Mode: A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:ABSolute:STOP <numeric_value>
This command defines the absolute limit at the stop frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter: -400 to +400 dBm
Example: "ESP:RANG5:LIM:ABS:STOP -40"
'Absolute limit at stop frequency in range 5 is -40 dBm
Characteristics: *RST value: -13dBm
SCPI: device-specific
Mode: A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:RELative:STARt <numeric_value>
This command defines the relative limit at the start frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter: -400 to +400 dBc
Example: "ESP:RANG5:LIM:REL:STAR -40"
'Relative limit at start frequency in range 5 is -40 dBc

Table of Contents