Operating Manual 1302.6163.12 - 03 6.181
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]FREQuency:SPAN 0 to f
max
In analyzer mode, this command defines the frequency span.
Example: "FREQ:SPAN 10MHz"
Characteristics: *RST value: f
max
with f
max
= maximum frequency
SCPI: conform
Mode: A
[SENSe<1|2>:]FREQuency:SPAN:FULL
In analyzer mode, this command sets the frequency span to its maximum.
Example: "FREQ:SPAN:FULL"
Characteristics: *RST value: -
SCPI: conform
Mode: A
This command is an event and therefore has no *RST value and no query.
[SENSe<1|2>:]FREQuency:STARt 0 to f
max
This command defines the start frequency of the overall scan in receiver mode. It defines the start
frequency of the sweep in analyzer mode..
Example: "FREQ:STAR 20MHz"
Characteristics: *RST value: 0
SCPI: conform
Mode: R, A-F
[SENSe<1|2>:]FREQuency:STOP 0 to f
max
This command defines the stop frequency of the overall scan in receiver mode. It defines the stop
frequency of the sweep in analyzer mode..
Example: "FREQ:STOP 20MHz"
Characteristics: *RST value: f
max
SCPI: conform
Mode: R, A-F
[SENSe<1|2>:]FREQuency:MODE CW | FIXed | SWEep | SCAN | TDOMain
In analyzer mode, this command switches between frequency domain (SWEep) and time domain (CW
| FIXed).
In receiver mode, this command switches between frequency scan (SCAN) and time domain scan
(TDOMain).
For CW and FIXed, the frequency setting is made using the FREQuency:CENTer command. For
SWEep, the setting is made using the FREQuency:STARt, STOP, CENTer and SPAN commands.
For SCAN, the frequency setting is made using the [SENSe<1|2>:]SCAN<1 to 10>:STARt and
[SENSe<1|2>:]SCAN<1 to 10>:STOP commands.