Operating Manual 1302.6163.12 - 03 6.189
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]LIST:RANGe<1…20>:COUNt?
This command indicates the number of defined ranges of the spurious measurement.
The numeric suffix <1 to 20> with RANGe is not relevant for this command.
Example: "LIST:RANG:COUN?"
'Indicates the number of ranges
Characteristics: *RST value: --
SCPI: device-specific
Mode: A
[SENSe<1|2>:]LIST:RANGe<1…20>:DELete
This command deletes a range.
Example: "LIST:RANG2:DEL"
'Deletes range 2
Characteristics: *RST value: --
SCPI: device-specific
Mode: A
[SENSe<1|2>:]LIST:RANGe<1…20>:DETector APEak | NEGative | POSitive | SAMPle | RMS |
AVERage
This command sets the detector in the spurious measurement.
Example: "LIST:RANG2:DET APE"
'Autopeak detector
Characteristics: *RST value: POS
SCPI: device-specific
Mode: A
[SENSe<1|2>:]LIST:RANGe<1…20>:FILTer:TYPE NORMal | CHANnel | RRC | P5 | P5D | PULSe
This command sets the filter in the spurious measurement. For details on filter type refer to chapter 4,
section “Filter Types” on page 4.117.
Example: "LIST:RANG2:FILT:TYPE RRC"
'RRC filter
Characteristics: *RST value: NORMal
SCPI: device-specific
Mode: A
[SENSe<1|2>:]LIST:RANGe<1…20>[:FREQuency]:STARt <numeric_value>
This command sets the start frequency of a range in the spurious measurement.
Example: "LIST:RANG2:STAR 1GHZ"
'Start frequency of range 2 to 1 GHz
Characteristics: *RST value: --
SCPI: device-specific
Mode: A

Table of Contents