Operating Manual 1302.6163.12 - 03 6.261
R&S ESU Remote Control – Description of Commands
TRIGger Subsystem
TRIGger<1|2>[:SEQuence]:LEVel:VIDeo 0 to 100PCT
This command sets the level of the video trigger source.
Example: "TRIG:LEV:VID 50PCT"
Characteristics: *RST value: 50 PCT
SCPI: device-specific
Mode: R, A
TRIGger<1|2>[:SEQuence]:HOLDoff -100 to 100s
This command defines the length of the trigger delay.
A negative delay time (pretrigger) can be set in the time domain (span = 0 Hz) only.
Example: "TRIG:HOLD 500us"
Characteristics: *RST value: 0 s
SCPI: conform
Mode: R, A
TRIGger<1|2>[:SEQuence]:HOLDoff:ADJust:AUTO ON|OFF
This command defines whether the group delay of the bandwidth filters for the external or IF power
trigger is compensated (ON) or not (OFF). If a burst signal is analyzed in the time domain mode, the
rising edge will remain at the same time position if compensation is on and the bandwidth changes.
Example: "TRIG:HOLD:ADJ:AUTO ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A
TRIGger<1|2>[:SEQuence]:SLOPe POSitive | NEGative
This command selects the slope of the trigger signal. The measurement starts on a positive or negative
slope of the trigger signal. The selected trigger slope applies to all trigger signal sources.
Example: "TRIG:SLOP NEG"
Characteristics: *RST value: POSitive
SCPI: conform
Characteristics: SCPI: device-specific
Mode: R, A