R&S ESU Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
6.324 Operating Manual 1302.6163.12 - 03
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:CENTer 0 to f
max
frequency ranges are
different for R&S ESU
and FSE
FSE [SENSe<1|2>:]FREQuency:CENTer:LINK STARt | STOP | SPAN not available for
R&S ESU
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:CENTer:STEP 0 to f
max
frequency ranges are
different for R&S ESU
and FSE
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:CENTer:STEP:LINK:
FACTor
1 to 100 PCT
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:MODE CW|FIXed | SWEep
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:OFFSet <numeric_value>
FSET [SENSe<1|2>:]FREQuency:RANGe 2 GHz | 22 GHz not available for
R&S ESU
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:SPAN 0 to f
max
frequency ranges are
different for R&S ESU
and FSE
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:SPAN:FULL
FSE [SENSe<1|2>:]FREQuency:SPAN:LINK CENTer | STOP | SPAN not available for
R&S ESU
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:STARt 0 to f
max
frequency ranges are
different for R&S ESU
and FSE
FSET [SENSe<1|2>:]FREQuency:STARt:FLINe[:STATe] ON | OFF not available for
R&S ESU. Replaced by
CALC:MARK:FUNC:
SLIMits
FSE [SENSe<1|2>:]FREQuency:STARt:LINK CENTer | STOP | SPAN not available for
R&S ESU
R&S
ESU
FSE [SENSe<1|2>:]FREQuency:STOP 0 to f
max
frequency ranges are
different for R&S ESU
and FSE
FSET [SENSe<1|2>:]FREQuency:STOP:FLINe[:STATe] ON | OFF not available for
R&S ESU; replaced by
CALC:MARK:FUNC:
SLIMits.
FSE [SENSe<1|2>:]FREQuency:STOP:LINK CENTer | STARt | SPAN not available for
R&S ESU
FSET/
ESIB
[SENSe<1|2>:]FREQuency[:CW|:FIXed] f
min
to f
max
not available for
R&S ESU
FSET/
ESIB
[SENSe<1|2>:]FREQuency[:CW|:FIXed]:STEP f
min
to f
max
not available for
R&S ESU
R&S
ESU
[SENSe<1|2>:]LIST:POWer:RESult? new function for
R&S ESU
Devices Command (Sheet 36 of 47) Parameter Notes