R&S ESU Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
6.334 Operating Manual 1302.6163.12 - 03
R&S
ESU
TRACe:IQ:SET <filter type>,
<rbw>,
<sample rate>,
<trigger source>,
<trigger slope>,
<pretrigger samples>,
<# of samples>
new function for
R&S ESU
R&S
ESU
TRACe:IQ:SRATe 16kHz to 32MHz new function for
R&S ESU
R&S
ESU
TRACe:IQ[:STATe] ON | OFF new function for
R&S ESU
R&S
ESU
FSE TRIGger<1|2>[:SEQuence]:HOLDoff 0 to 100s
FSE TRIGger<1|2>[:SEQuence]:LEVel:AF -120 to +120PCT not available for
R&S ESU
FSE TRIGger<1|2>[:SEQuence]:LEVel:VIDeo 0 to 100PCT not available for
R&S ESU; replaced by
TRIGger:SEQuence:
SOURce:VIDeo
FSE TRIGger<1|2>[:SEQuence]:LEVel[:EXTernal] -5.0 to +5.0V not available for
R&S ESU
R&S
ESU
TRIGger<1|2>[:SEQuence]:LEVel:IFPower <numeric_value> new command for
R&S ESU
R&S
ESU
FSE TRIGger<1|2>[:SEQuence]:SLOPe POSitive | NEGative
R&S
ESU
TRIGger<1|2>[:SEQuence]:SOURce IMMediate | LINE |
EXTernal | VIDeo |
IFPower
FSE/
ESIB
TRIGger<1|2>[:SEQuence]:SOURce IMMediate | LINE |
EXTernal | VIDeo |
RFPower | TV | AF
FSIQ TRIGger<1|2>[:SEQuence]:SOURce IMMediate | LINE |
EXTernal | VIDeo |
RFPower | AF
FSET TRIGger<1|2>[:SEQuence]:SOURce IMMediate | LINE |
EXTernal | VIDeo
R&S
ESU
FSE TRIGger<1|2>[:SEQuence]:SOURce:VIDeo:
FORMat:LPFrame
525 | 625 requires option B6 on
FSP
R&S
ESU
FSE TRIGger<1|2>[:SEQuence]:SOURce:VIDeo:
FIELd:SELect
ALL|ODD|EVEN requires option B6 on
FSP
R&S
ESU
FSE TRIGger<1|2>[:SEQuence]:SOURce:VIDeo:LINE:
NUMBer
<numeric_value> requires option B6 on
FSP
R&S
ESU
FSE TRIGger<1|2>[:SEQuence]:SOURce:VIDeo:
SSIGnal:POLarity
NEGative | POSitive requires option B6 on
FSP