Operating concepts
R&S
®
NRT-Z14/44
25Getting Started 1443.8632.02 ─ 01
1 = signal source
2 = R&S NRTZxx directional power sensor
3 = load
4 = computer with installed R&S Virtual NRT
5 = R&S NRTZ5 USB interface adapter
To start a measurement
1. Connect the R&S NRT-Zxx to the computer as shown in "Setup" on page 24.
For detailed descriptions, refer to:
Chapter 3.4, "Connecting to source and load", on page 14
Chapter 3.5.1, "Computer", on page 16
2. Switch on the test signal of the signal source.
3. Start R&S Virtual NRT.
A few seconds after the user interface is displayed, when the pointers of the
analog indication start wriggling, the power sensor is ready for operation. For
further instructions, see the R&S Virtual NRT help.
If communication with the power sensor could not be established, the display
is dimmed and the automatic configuration of the interface failed. Proceed
with "To establish the communication" on page 26.
R&S Virtual NRT