EasyManua.ls Logo

R&S RTO6 - Page 1082

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Jitter analysis and clock data recovery
R&S
®
RTO6
1082User Manual 1801.6687.02 ─ 05
Analyzed source
Selects the source for the advanced jitter analysis.
Remote command:
ADVJitter<m>:SIGNal:
SOURce on page 2486
Serial standard
For some serial standards, the bit rate is known to the instrument. If you select the
standard, the nominal bit rate for the advanced jitter analysis is set automatically.
The serial standards are available only if the instrument is able to support its maximum
nominal bit rate.
Serial
standard
Nom. bit
rate
Bandwidth Rel. bandwidth Order Damping
USB 2.0 480 Mbps 1667 287.947 kHz 1st -
USB 3.1
Gen 1
5 Gbps 500 10 MHz 2nd 0.7
Displayport
1.1
1.296
Gbps
1667 777.44 kHz 1st -
Displayport
1.2
4.32 Gbps 1667 2.59 MHz 1st -
Fibre
Channel 1x
1 Gbps 1667 599.88 kHz 1st -
Fibre
Channel 2x
2 Gbps 1667 1.2 MHz 1st -
Advanced jitter and noise (option R&S RT
O6- K133/K134)

Table of Contents

Other manuals for R&S RTO6

Related product manuals