1249User Manual 1801.6687.02 ─ 05
Bit No. Meaning
6
FREQuency
This bit is set if there is anything wrong with the frequency of the local oscillator or the reference
frequency (see "STA
Tus:QUEStionable:FREQuency register" on page 1250).
7
DEEMbedding
The bit is set if a warning or an error occurs during deembedding (see "STATus:QUEStiona-
ble:DEEMbedding register" on page 1251).
8
NOALigndata
This bit is set if no alignment data is available - the instrument is uncalibrated.
9
LIMit
This bit is set if a limit value is violated (see "STATus:QUEStionable:LIMit, STATus:QUEStiona-
ble:MARGin registers" on page 1251).
10
MARGin
This bit is set if a margin value is violated, and the limit value is not violated (see "STA-
Tus:QUEStionable:LIMit, STATus:QUEStionable:MARGin registers" on page 1251).
11 LAMP (Low AMPlitude)
This bit is set if the magnitude of the signal is too low to get reliable measurement results. See
"STA
Tus:QUEStionable:LAMP register" on page 1252.
12
MASK
This bit is set if a mask value is violated (see "STATus:QUEStionable:MASK register"
on page 1252
13 Not used
14 ZVCO: This bit is set if a channel overload occurs at one or more input channels of the R&S RT
-
ZVC multi-channel power probe. See "STATus:QUEStionable:ZVCO register" on page 1252.
15 This bit is always 0.
STATus:QUEStionable:COVerload register
This register contains all information about overload of the channels. The bit is set if
the assigned channel is overloaded, or if an overload risk occurred (overload warning).
Table 22-4: Bits in the STATus:QUEStionable:COVerload register
Bit No. Meaning
0 Overload on CHANnel1
1 Overload on CHANnel2
2 Overload on CHANnel3
3 Overload on CHANnel4
4 Overload warning for CHANnel1
5 Overload warning for CHANnel2
6 Overload warning for CHANnel3
7 Overload warning for CHANnel4