Remote control commands
R&S
®
RTO6
1367User Manual 1801.6687.02 ─ 05
ZVC:Z<m>:V<n>[:STATe]..............................................................................................
1375
ZVC:Z<m>:ID:
NAME?.................................................................................................. 1375
ZVC:Z<m>:ID:PARTnumber?........................................................................................ 1376
ZVC:Z<m>:ID:SRNumber?............................................................................................1376
ZVC:Z<m>:ID:SWVersion?............................................................................................1376
ZVC:Z<m>:I<n>:ZERComp:DETect................................................................................1376
ZVC:Z<m>:I<n>:ZERComp:USE................................................................................... 1377
ZVC:Z<m>:I<n>:ZERComp:STATe?...............................................................................1377
ZVC:BANDwidth <Bandwidth>
Sets the bandwidth limit of the probe. The bandwidth specifies the maximum frequency
at which a purely sinusoidal signal is still transferred at 89 % (0.1 dB) of its amplitude.
Parameters:
<Bandwidth> Range: 5000 to 1E+6
Increment: 5000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
ZVC:TYPE <DecimationMode>
Sets the decimation mode for the R&S RT
-ZVC probe. Decimation reduces the data
stream of the ADC to a stream of waveform points with lower sample rate and a less
precise time resolution.
Parameters:
<DecimationMode> SAMPle | PDETect | HRESolution
*RST: SAMPle
Usage: Asynchronous command
ZVC:Z<m>:I<n>:BANDwidth?
Queries the bandwidth of the current channel. You can set the probe bandwidth with
ZVC:BANDwidth.
The bandwidth of some current channels is restricted to 300KHz due to their vertical
settings.
Suffix:
<m>
.
1..2
<n> 1..4
Return values:
<Bandwidth>
Range: 5000 to 1E+6
Increment: 5000
*RST: 1E+6
Default unit: Hz