Remote control commands
R&S
®
RTO6
1538User Manual 1801.6687.02 ─ 05
MEASurement<m>:VERTical:SCALe <V
erticalScale>
Defines the vertical scaling per division, so that the scaling can be adapted automati-
cally during the long term measurement period.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<V
erticalScale> Range: 1E-12 to 100E+12
Increment: 10E-6
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
MEASurement<m>:LTMeas[:STATe] <ShowDiagram>
Enables long term measurement for a defined number of measurement points (see
MEASurement<m>:
LTMeas:
COUNt on page 1538) or a specified time
(seeMEASurement<m>:LTMeas:TIME on page 1539).
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<ShowDiagram>
ON | OFF
*RST: OFF
Usage: Asynchronous command
MEASurement<m>:LTMeas:COUNt <MeasCnt>
Defines the total number of points to be measured during the long term measurement.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<MeasCnt>
Range: 1000 to 200000
Increment: 10
*RST: 1000
Usage: Asynchronous command
Automatic measurements