Remote control commands
R&S
®
RTO6
1734User Manual 1801.6687.02 ─ 05
Parameters:
<DataPosOperator> EQUal | GETHan | INRange | RANGe
INRange = RANGe
*RST: GETHan
Usage: Asynchronous command
TRIGger<m>:UART:DPOSition <DataPosition>
Sets the number of words before the first word of interest. These offset words are
ignored.
Parameters:
<DataPosition>
Number of words
Range: 0 to 32767
Increment: 1
*RST: 0
Usage: Asynchronous command
TRIGger<m>:UART:DPTO <DataPositionT
o>
Defines the last word of interest, if
TRIGger<m>:UART:DPOPerator defines a posi-
tion range.
Parameters:
<DataPositionTo> Range: 0 to 32767
Increment: 1
*RST: 0
Usage: Asynchronous command
TRIGger<m>:UART:FCONdition <DataOperator>
Selects the operator for the data pattern (TRIGger<m>:UART:DATA).
Parameters:
<DataOperator>
EQUal | NEQual
*RST: EQUal
TRIGger<m>:UART:DATA <Data>
Specifies the data pattern to be found on the specified trigger source, in binary or hex
format. Enter the words in msb first bit order.
Parameters:
<Data> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
Usage: Asynchronous command