Parameters:
<Bitrate> Range: 100 to 5E+6
Increment: 1
*RST: 100E+3
Default unit: bps
BUS<m>:CAN:FDATa:ENABle <CANFDEnabled>
Enables the CAN FD protocol configuration.
The setting is available in CAN FD option.
Suffix:
<m>
.
1..4
Parameters:
<CANFDEnabled> ON | OFF
*RST: ON
BUS<m>:CAN:FDATa:DBITrate <FlexDatBitrate>
Sets the bit rate of the data phase. The data rate can be higher than the arbitration
rate, but it is uniform and fixed for a given CAN FD bus.
The setting is available in CAN FD option.
Suffix:
<m>
.
1..4
Parameters:
<FlexDatBitrate> Range: 100 to 15E+6
Increment: 1
*RST: 1E+6
Default unit: bps
BUS<m>:CAN:FDATa:SAMPlepoint <FlexDatSmpPt>
BUS<m>:CAN:SAMPlepoint
<SamplePoint>
Sets the position of the sample point within the bit in percent of the nominal bit time.
Alternatively, you can set the sample point with BUS<m>:CAN:T1Segment and
BUS<m>:CAN:T2Segment.
For CAN FD signals, BUS<m>:CAN:SAMPlepoint defines the synchronization of the
arbitration phase, and BUS<m>:CAN:FDATa:SAMPlepoint defines the synchroniza-
tion of the data phase.
Suffix:
<m>
.
1..4
Selects the serial bus.