Remote control commands
R&S
®
RTO6
1869User Manual 1801.6687.02 ─ 05
BUS<m>:FLXRay:POLarity <Polarity>
Selects the wire on which the bus signal is measured in case of SINGle source type.
The setting af
fects the digitization of the signal.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Polarity> BPLus | BMINus
*RST
: BPLus
BUS<m>:FLXRay:BITRate <Bitrate>
Selects the number of transmitted bits per second.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Bitrate> M10 | M5 | M2_5
10, 5, or 2.5 Mbit/s.
The return value of 2.5 Mbit/s is M25.
*RST: M10
Example:
BUS:FLXRay:BITRate M2_5
BUS:FLXRay:
BITRate?
M25
BUS<m>:FLXRay:CHTYpe <Channel>
Selects the channel on which the signal is measured. The setting is considered in the
calculation of the frame CRC.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Channel> CHA | CHB
Channel A or channel B
*RST: CHA
BUS<m>:FLXRay:SEHB <SeparateHdrBts>
The command affects the decoding and its display
. If ON, the leading five indicator bits
of the header are decoded as five single bits. Otherwise, the indicator bits are shown
as one word with word length five bit.
Protocols