EasyManua.ls Logo

R&S RTO6 - Page 1913

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1913User Manual 1801.6687.02 ─ 05
Return values:
<State> OK | UNDF
UNDF: Undefined
*RST: OK
Usage: Query only
BUS<m>:SENT:FRAMe<n>:NIBBle<o>:VALue?
Returns the value of the specified nibble.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
<o> *
Selects the nibble number.
Return values:
<V
alue> To set the value format, use FORMat:BPATtern
.
The values below range, increment and reset are decimal
values.
Range: 0 to 15
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:SENT:FRAMe<n>:PAPTicks?
Returns the number of the pulse pause clock ticks.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<PausePulseTicks> Range: 12 to 768
Increment: 1
*RST: 12
Usage: Query only
BUS<m>:SENT:FRAMe<n>:SCOM?
Returns the value of the status/communication pulse.
Protocols