Remote control commands
R&S
®
RTO6
1959User Manual 1801.6687.02 ─ 05
BUS<m>:MILStd:POLarity <Polarity>
Selects the wire on which the bus signal is measured.
Suffix:
<m>
.
1..4
Parameters:
<Polarity> NORMal | INVerted
*RST: NORMal
BUS<m>:MILStd:PRESet <Preset>
Sets the default threshold voltage.
Suffix:
<m>
.
1..4
Parameters:
<Preset> V05 | V2 | V5 | V7 | MANual
*RST: V5
BUS<m>:MILStd:THReshold:HIGH <ThresholdHigh>
Sets the lower threshold level of the signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdHigh> Range: 0 to 14
Increment: 0.1
*RST: 5
Default unit: V
BUS<m>:MILStd:THReshold:LOW <ThresholdLow>
Sets the lower threshold level of the signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdLow> Range: -14 to 0
Increment: 0.1
*RST: -5
Default unit: V