Remote control commands
R&S
®
RTO6
2063User Manual 1801.6687.02 ─ 05
Slave address /MID TRIGger<m>:RFFE:SADD:CONDition
TRIGger<m>:
RFFE:
SADD:MIN
TRIGger<m>:RFFE:SADD:MAX
Register address TRIGger<m>:RFFE:ADDRess:CONDition
TRIGger<m>:
RFFE:
ADDRess:MIN
TRIGger<m>:RFFE:ADDRess:MAX
Data pattern TRIGger<m>:RFFE:DATA:DCON
TRIGger<m>:
RFFE:
DATA:DMIN
TRIGger<m>:RFFE:DATA:DMAX
Mask address TRIGger<m>:RFFE:MASK:CONDition
TRIGger<m>:
RFFE:
MASK:MAX
TRIGger<m>:RFFE:MASK:MIN
Byte count TRIGger<m>:RFFE:BCOunt:CONDition
TRIGger<m>:
RFFE:
BCOunt:MIN
TRIGger<m>:RFFE:BCOunt:MAX
Data index TRIGger<m>:RFFE:DATA:ICONdition
TRIGger<m>:
RFFE:
DATA:IMIN
TRIGger<m>:RFFE:DATA:IMAX
TRIGger<m>:RFFE:ADDRess:CONDition <AddressOperator>
Sets the operator to trigger on a specific address pattern or a range.
Parameters:
<AddressOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one pattern to
be set with
TRIGger<m>:RFFE:ADDRess:MIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:RFFE:ADDRess:MIN and
TRIGger<m>:RFFE:ADDRess:MAX.
*RST: EQUal
TRIGger<m>:RFFE:ADDRess:MIN <AddressMin>
Specifies a pattern, or sets the start value of a pattern range.
Parameters:
<AddressMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (do not care).