Remote control commands
R&S
®
RTO6
2067User Manual 1801.6687.02 ─ 05
TRIGger<m>:RFFE:SADD:MIN <SIDMin>
Specifies a pattern, or sets the start value of a pattern range.
Parameters:
<SIDMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:RFFE:SADD:MAX <SIDMax>
Sets the end value of a range, if the slave address/MID condition is set to INRange or
OORange. See SEARch:TRIGger:
RFFE:SADD:CONDition.
Parameters:
<SIDMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:RFFE:MASK:CONDition <MaskOperator>
Sets the operator to trigger on a mask pattern or a range.
Parameters:
<MaskOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one pattern to
be set with
TRIGger<m>:RFFE:MASK:MIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:RFFE:MASK:MIN and
TRIGger<m>:RFFE:MASK:MAX.
*RST: EQUal
TRIGger<m>:RFFE:MASK:MAX <MaskMax>
Sets the end value of a range, if themask condition is set to INRange or
OORange.
See TRIGger<m>:RFFE:MASK:CONDition.
Parameters:
<MaskMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:RFFE:MASK:MIN <MaskMin>
Specifies a mask pattern, or sets the start value of a mask pattern range.