Remote control commands
R&S
®
RTO6
2157User Manual 1801.6687.02 ─ 05
TRIGger<m>:USB:FMAX <FrameNumberT
o>
Sets the the end value of a frame number range if TRIGger<m>:
USB:FCONdition is
set to INRange or OORange.
Parameters:
<FrameNumberTo>
TRIGger<m>:USB:HAND <HandshakeType>
Sets the trigger to one out of four dif
ferent handshake types: ACK, NAK, STALl, or
NYET.
Parameters:
<HandshakeType> ACK | NAK | STALl | NYET
ACK
Sets the trigger to the ACK field (acknowledgment of error-free
data packet).
NAK
Sets the trigger to the NAK field (non-acknowledgment, no suc-
cessful data transmission).
ST
ALl
Sets the trigger to the ST
ALL field (endpoint is halted or a con-
trol pipe request is not supported).
NYET
Sets the trigger to the NYET field (no response yet from
receiver).
*RST: ACK
TRIGger<m>:USB:PATT <PayloadMuster>
Specifies the payload data pattern that is to be searched.
Parameters:
<PayloadMuster>
TRIGger<m>:USB:PCONdition <OperatorPort>
Sets the operator to set a specific port number or a port number range. The port num-
ber values are set with TRIGger<m>:USB:PMIN
and TRIGger<m>:USB:PMAX.
Parameters:
<OperatorPort> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one port number
to be set using
TRIGger<m>:USB:PMIN.