Triggers
R&S
®
RTO6
226User Manual 1801.6687.02 ─ 05
"Polarity"
Sets the polarity of the trigger out pulse, that is the direction of the
first pulse edge.
"Delay"
Only available if "Trigger out" is set to "On trigger". The setting
defines the delay of the first pulse edge to the trigger point. The mini-
mum delay is 800 ns.
"Pulse length"
Sets the length of the trigger out pulse.
Remote command:
TRIGger<m>:
OUT:POLarity
on page 1430
TRIGger<m>:OUT:PLENgth on page 1431
TRIGger<m>:OUT:DELay on page 1431
6.7 Qualification
Access: "Menu" > "Trigger" > "Qualify" tab
Qualification