Remote control commands
R&S
®
RTO6
2291User Manual 1801.6687.02 ─ 05
Parameters:
<PACPFunIDMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page
1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY
:PFID:MAX <PACPFunIDMax>
Sets the end value of the P
ACP Function ID range if TRIGger<m>:MPHY:PFID:
CONDition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<PACPFunIDMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:PRPLength <Width>
Sets the width for the width trigger based upon the data rate.
Suffix:
<m>
.
1..3
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger<m>:MPHY:ERRor:LCMD <LCCError>
Enables/disables trigger on LCC error.
Suffix:
<m>
.
1..3
Parameters:
<LCCError>
ON | OFF
*RST: ON
TRIGger<m>:MPHY:ERRor:LENGth <ErrorLength>
Enables/disables trigger on length error.