Remote control commands
R&S
®
RTO6
2365User Manual 1801.6687.02 ─ 05
Return values:
<SequenceNumber> Range: 0 to 4095
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:PCIE:RESult:PACKet<n>:STARt?
Returns the start time of the specified packet.
Suffix:
<m>
.
1..4
<n> *
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:PCIE:RESult:PACKet<n>:STATe?
Returns the state of the specified packet.
Suffix:
<m>
.
1..4
<n> *
Return values:
<State> OK | MALF | CRC16 | ECRC | LCRC | LEN | POE | MEE |
UNCorrelated | FCRC |
FPERror | UNKNown | OUTRange |
INComplete | SYNCerror
MALF
Mal formated packet
CRC16
CRC16 error
ECRC
ECRC error
LCRC
LCRC error
LEN
Length error
POE
TLP prefix error
MEE
Max end-end TLP prefix error
Protocols