Remote control commands
R&S
®
RTO6
2427User Manual 1801.6687.02 ─ 05
<n> 0..15
Selects the bit of the bus word. Each bit corresponds to a digital
channel.
Parameters:
<Deskew> Range: -200E-9 to 200E-9
Increment: 200E-12
*RST: 0
Default unit: s
BUS<m>:PARallel:DESoffset <DeskewOffset>
Sets a general delay for all digital channels. The resulting deskew of a digital channel
is the sum of this general value and the individual value set with BUS<m>:
PARallel:
BIT<n>:DESKew.
Suffix:
<m>
.
1..4
Selects the parallel bus.
Parameters:
<DeskewOffset> Range: -200E-9 to 200E-9
Increment: 200E-12
*RST: 0
Default unit: s
BUS<m>:PARallel:BIT<n>:LABel <Label>
Sets a name for the indicated digital channel. The name is displayed in the diagram.
For the parallel bus 1, you can also use DIGital<m>:
LABel.
Suffix:
<m>
.
1..4
Selects the parallel bus.
<n> 0..15
Selects the bit of the bus word. Each bit corresponds to a digital
channel.
Parameters:
<Label> String containing the channel name
BUS<m>:PARallel:DATA:FORMat <DataFormat>
Sets the number format for decoded data values on the display for the indicated paral-
lel bus.
It sets also the number representation for remote data transfer with BUS<m>:
PARallel:DATA[:VALues]?.
Suffix:
<m>
.
1..4
Mixed signal option (MSO, R&S RT
O6-B1)