WGENerator<m>:MODulation:AM:FREQuency <Frequency>
Sets the frequency of the modulation waveform for amplitude modulation, if
WGENerator<m>:
MODulation:
TYPE is set to AM.
Suffix:
<m>
.
1..2
Parameters:
<Frequency> Range: 1E-3 to 1E+6
Increment: 1
*RST: 1000
Default unit: Hz
WGENerator<m>:MODulation:AM:SYMMetry <RampSymmetry>
Sets the symmetry, the percentage of time the ramp modulation waveform is rising, for
amplitude modulation, if WGENerator<m>:
MODulation:TYPE is set to AM.
Suffix:
<m>
.
1..2
Parameters:
<RampSymmetry> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
WGENerator<m>:MODulation:CARRier:FREQuency <CarrierFreq>
Sets the frequency of the carrier signal for a modulation waveform.
Suffix:
<m>
.
1..2
Parameters:
<CarrierFreq> Range: 1E-3 to 100E+6
Increment: 10
*RST: 1E+6
Default unit: Hz
WGENerator<m>:MODulation:CARRier:PERiod <CarrierPeriod>
Sets the period of the carrier signal for a modulation waveform.
Suffix:
<m>
.
1..2
Waveform generator (option R&S RT
O6-B6)