Remote control commands
R&S
®
RTO6
2469User Manual 1801.6687.02 ─ 05
Parameters:
<UseCalibrationDat> ON | OFF
*RST: ON
Usage: Asynchronous command
TDRT:BW <Bandwidth>
Sets the bandwidth of the DUT.
Parameters:
<Bandwidth>
Range: 2E+9 to 16E+9
Increment: 10
*RST: 4E+9
Default unit: Hz
Usage: Asynchronous command
TDRT:DOMain <OutputDomain>
Sets the domain of the x-axis: time or distance (in m or inch). If you select distance,
enter also the permittivity with TDRT:PERMittivity.
Parameters:
<OutputDomain> TIME | MDIStance | INDistance
MDIStance = distance in m
IDIStance = distance in inch
*RST: TIME
Usage: Asynchronous command
TDRT:PERMittivity <Permittivity>
Sets the permittivity "ε_r" of the used DUT.
Parameters:
<Permittivity>
Range: 1 to 100
Increment: 0.1
*RST: 3.3
Usage: Asynchronous command
TDRT:SIGNal<m>STATe <State>
Enables the indicated TDR/TDT signal.
Suffix:
<m>
.
1..3
1 = impedance Z (result of TDR measurement)
2 = reflection coefficient ρ (result of TDR measurement)
3 = step response h (result of TDT measurement)
Pulse Source and TDR/TDT Analysis (options R&S RT
O6-B7/K130)