2490User Manual 1801.6687.02 ─ 05
ADVJitter<m>:SIGNal:GATE:RANGe:ABSolute:STOP <StopV
alue>
Sets the stop time of the gate area in percent, if
ADVJitter<m>:SIGNal:GATE:
RANGe:MODE is set to ABS.
Suffix:
<m>
.
1..2
Parameters:
<StopValue> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
ADVJitter<m>:SIGNal:GATE:RANGe:MODE <ModeT
ype>
Selects if the gate area is defined in relative values (percent) or absolute.
Suffix:
<m>
.
1..2
Parameters:
<ModeT
ype> ABS | REL
*RST: ABS
ADVJitter<m>:SIGNal:GATE:RANGe:RELative:STARt <StartValue>
Sets the relative start of the gate area in percent, if
ADVJitter<m>:SIGNal:GATE:
RANGe:MODE is set to REL.
Suffix:
<m>
.
1..2
Parameters:
<StartValue> Range: 0 to 100
Increment: 0.1
*RST: 0
Default unit: %
ADVJitter<m>:SIGNal:GATE:RANGe:RELative:STOP <StopValue>
Sets the relative stop of the gate area in percent, if ADVJitter<m>:
SIGNal:GATE:
RANGe:MODE is set to REL.
Suffix:
<m>
.
1..2
Parameters:
<StopValue> Range: 0 to 100
Increment: 0.1
*RST: 100
Default unit: %