Mask testing
R&S
®
RTO6
406User Manual 1801.6687.02 ─ 05
●
"Acquisitions": Considers the number of acquisitions that contain at least one sam-
ple hit. How many samples hit the mask in that acquisition is not relevant.
"V
iolation tolerance" sets the number of tolerable sample hits or acquisition hits.
A test has failed if the number of sample hits or acquisition hits exceeds the limit of vio-
lation tolerance hits.
Remote command:
MTESt:CONDition on page 1584
MTESt:TOLerance on page 1584
Action on mask
The settings in this dialog define what happens when the mask test has failed or when
it has passed successfully.
Most actions can be initiated either on failure or on success:
●
On violation
The action is initiated when the fail criteria is fulfilled.
●
On successful completion
The action is initiated when the [Single] acquisition has finished and the fail criteria
is not fulfilled - the fail condition and violation tolerance limit have not been
reached.
Beep ← Action on mask
Generates a beep sound.
Remote command:
MTESt:
ONViolation:
BEEP on page 1594
Stop acq. ← Action on mask
Stops the waveform acquisition on mask violation.
Remote command:
MTESt:ONViolation:STOP on page 1595
Save screenshot ← Action on mask
Saves a screenshot including the mask test results according to settings in "Menu" >
"Save/Recall" > "Save" tab > "Screenshot".
Mask test settings