Protocol analysis
R&S
®
RTO6
763User Manual 1801.6687.02 ─ 05
See: "Address" on page
766
Set a data pattern condition to trigger on data patterns expected in the message.
See:
"Data" on page 767 and "Index" on page 767
Extended Register Read ← Type
Triggers on "Extended Register Read" command sequences. This sequence sends the
byte count of data frames in the command payload, followed by one address frame
with the address of the first extended register, and reads back up to 16 data frames.
You can refine the trigger condition using the same settings as for the "Extended Reg-
ister Write" command sequence, see "Extended Register Write" on page 762.
Extended Register Write Long ← Type
Triggers on "Extended Register Write" command sequences. This sequence sends the
byte count of data frames in the command payload, followed by 2 address frames with
the address of the first extended register, and up to 8 data frames.
You can refine the trigger condition using the same settings as for the "Extended Reg-
ister Write" command sequence, see "Extended Register Write" on page 762.
Extended Register Read Long ← Type
Triggers on "Extended Register Write" command sequences. This sequence sends the
byte count of data frames in the command payload, followed by 2 address frames with
the address of the first extended register, and reads back up to 8 data frames.
You can refine the trigger condition using the same settings as for the "Extended Reg-
ister Write" command sequence, see "Extended Register Write" on page 762.
Interrupt Summary and Notification ← Type
Available only for "Version"= "2.0"/"2.1".
Triggers on "Interrupt Summary and Notification" command sequence.
You can refine the trigger condition:
Set a slave address condition to trigger only on command sequences that are sent
to the specified MIDs.
See: "Slave Addr. /MID " on page 766
Set a bit count condition to trigger only on command sequences that send the
specified number of data frames.
See: "Byte count" on page 766
Set a data pattern condition to trigger on data patterns expected in the message.
See: "Data" on page 767
Masked Write ← Type
Available only for "Version"= "2.1".
Triggers on "Masked Write" command sequences. This sequence sends the register
address in the command payload, and sends one mask and one data frame.
You can refine the trigger condition:
Set a slave address condition to trigger only on command sequences that are sent
to the specified MIDs.
See: "Slave Addr. /MID " on page 766.
Set a register address condition to trigger only on command sequences that write
data to the specified register.