Command reference
R&S
®
ZNA
1400User Manual 1178.6462.02 ─ 20
Parameters:
<Value> *RST: 1 ms
Default unit: s
Options: R&S ZNA-K28
Manual operation: See "Wait Time per Point" on page 491
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE <Boolean>
Enables/disables user-defined wait times per measurement point for the controlled tim-
ing mode (INITiate<Ch>:CTIMing ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> OFF (0)
Wait time is calculated by the analyzer firmware
ON (1)
Wait time can be specified manually using [SENSe<Ch>:
]SWEep:CTIMing:STIMe
Options: R&S ZNA-K28
Manual operation: See "Sweep Time Mode" on page 490
[SENSe<Ch>:]SWEep:DETector:TIME <DetectorTime>
Defines the observation time per sweep point if a detector other than the "Normal"
(SAMple) detector is used. The detector is selected together with the measured quan-
tity (CALCulate<Ch>:PARameter:MEASure or CALCulate<Ch>:PARameter:
SDEFine).
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime> Detector time
Range: 0 s to 3456000 s
*RST: 0.01 s
Default unit: s
Example:
*RST; :CALC:PAR:MEAS 'TRC1', 'a1D1AVG'
Select the wave quantity a
1
for the default trace and activate the
AVG detector.
SWE:DET:TIME 1
Specify an observation time of 1 s at each sweep point.
Manual operation: See "Detector" on page 309
SCPI command reference