Command reference
R&S
®
ZNA
1536User Manual 1178.6462.02 ─ 20
Example:
*RST; SENS1:FREQ:STAR 1 GHz; STOP 2 GHz
Select a stimulus range between 1 GHz and 2 GHz.
FREQ:MDEL:APER 1 MHz
Select a 1 MHz aperture.
FREQ:MDEL:UTON PORT, 3
Select NWA port 3 as a source port for the upper tone (no exter-
nal generator).
FREQ:MDEL:COMB ON
Use the internal combiner to generate the two-tone signal at port
1 (for instruments equipped with the necessary options).
...
(Perform mixer delay measurement, evaluate results.)
FREQ:MDELay:CONV OFF
Disable mixer delay measurement, switch back to normal (non
frequency-converting) mode.
Manual operation:
See "2 Tone Delta Freq." on page 372
[SENSe<Ch>:]FREQuency:MDELay:CDELay <const delay value for ref mixer>
Defines a constant mixer delay value, to be used as a reference for a mixer delay mea-
surement calibration. The setting takes effect after constant mixer delay is enabled;
see example below.
Suffix:
<Ch>
.
Channel number
Parameters:
<const delay value for
ref mixer>
Constant delay
Range: 0 to 1 ms
*RST: 0
Default unit: s
Example:
*RST; :FREQ:MDEL:CDEL 1 ns
Define a constant mixer delay value of 1 ns.
MMEM:LOAD:MDAD 1, CDEL
Use constant delay value for mixer delay measurement calibra-
tion.
Manual operation: See "Constant Cal Dly/Variable Cal Dly" on page 372
[SENSe<Ch>:]FREQuency:MDELay:CDMode <arg0>
Suffix:
<Ch>
.
Parameters:
<arg0>
Manual operation: See "Use Constant Cal Dly" on page 372
SCPI command reference