GUI reference
R&S
®
ZNA
336User Manual 1178.6462.02 ─ 20
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
CW Mode Spectrum
In CW spectrum mode, the intermodulation measurement is performed at fixed fre-
quency of the lower tone ("CW Frequency") and the upper tone ("CW Frequency" +
"Tone Distance"). The analyzer displays intermodulation products up to a selectable
order ("IM Order") around the lower and upper tone frequencies. The channel settings
differ from the swept intermodulation measurement, therefore a new channel is created
when the spectrum measurement is activated ("Add CW Mode").
The following example shows an intermodulation spectrum with the following settings:
●
CW Frequency: 2 GHz (lower tone, position of marker M1)
●
Tone Distance: 3 MHz (defines the upper tone, position of marker M2)
●
Max. IM Order: 9.
From left to right, the peaks correspond to the following intermodulation products:
– lower IM9
– lower IM7
– lower IM5 (marker M6)
– lower IM3 (marker M5)
– lower tone (marker M1)
– upper tone (marker M2)
– upper IM3 (marker M3)
– upper IM5 (marker M4)
– upper IM7
– upper IM9
Meas softtool